Properties | Details |
---|---|
SFDA Classification | Class IIb |
Purpose of Use | The premature baby incubator is designed to provide a safe and secure environment for premature babies until their full development. The appropriate incubator protects the baby's skin, with these special characteristics, the incubator protects premature babies until they are fully developed. |
Work Location & End-User | Place of work: neonatal intensive care and operating rooms End User: Medical staff, especially nursing |
Parts & Accessories | Temperature probe (one time use) probe cover • Air Purifier • Replacement for sleeves • Mattresses and covers |
Operation | A fan blows filtered ambient air over a heating element and a water container. Through a control valve additional oxygen can be supplied to the air. The moistened, heated and enriched air now flows into the above cabinet with the baby. One part of the air escapes from the cabinet through vent holes, another part gets back into the air processing. |
Common Problems | • Heating element malfunction • thermal breaker • a fan • brakes and doors |
Manufacturers | • Dräger • GE Healthcare • Hill-Room • Ohio Medical |
Sources | • Saudi Food and Drug Authority • Saudi Food and Drug Authority Document • Healthline • Tarry • Franks Hospital Workshop |